Dansensor® Lippke® 5000 EN User Guide 65
P/N 380419-C
05/2022
Display of test results
When a test terminates, the test result is displayed on the screen.
There are different possible test results depending on the type of test.
The various test results are described below.
Burst test
Leak test
PASS
825 mbar
FAIL
775 mbar
23 sec
FAIL
935 mbar
FAIL
No Burst
Package bursts within
the specified pressure
range
Package bursts below
the specified pressure
range.
Package bursts after
23 sec.
Package bursts above
the specified pressure
range
Package does not
burst
PASS
1 mbar
FAIL
13 mbar
23 sec
FAIL
Package Burst
Pressure loss in the
package is below the
specified leak limit
Pressure loss in the
package is above the
specified leak limit.
Limit is exceeded
after 23 sec.
Package bursts