EasyManua.ls Logo

Mocon AMETEK Dansensor Lippke 5000 - Page 82

Mocon AMETEK Dansensor Lippke 5000
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
80 User Guide EN Dansensor® Lippke® 5000
P/N 380419-C
05/2022
3017 Test time is lower than test
time upper limit
Incorrect parameter
settings when creating a
Creep2Fail Test Definition
Set appropriate values.
3107 Inlet pressure exceeds
allowed limit
Inlet pressure exceeds
max. limit
Reduce the inlet pressure
to an acceptable value.
3111 Initial package pressure
too high
The start pressure in the
package is higher or the
same as the test pressure.
Remove pressure from the
package or increase the
test pressure.
3112 Target pressure not
achieved
The package could not be
filled to the set pressure
Ensure that correct
package connection has
been selected.
Ensure that the package is
restrained, if needed.
Do not touch the package
when the device is
inflating the package.
Ensure that the entered
volume is correct.
If it is a porous package
cover the porous surface.
If it is a burst test reduce
the inflation rate.
For other tests reduce the
test speed.
3113 Overshoot more than XXX The pressure in the
package became higher
that the set point.
Ensure that the package is
restrained, if needed.
Do not touch the package
when the device is
inflating the package.
If it still overshoots, reduce
the test speed.
No. Message Problem Solution

Table of Contents

Related product manuals