Dansensor® Lippke® 5000 EN User Guide 81
P/N 380419-C
05/2022
3114 Regulation error The package could not be
filled with the settings
selected.
Ensure that correct
package connection has
been selected.
Ensure that the package is
restrained, if needed.
Do not touch the package
when the device is
inflating the package.
Ensure that the entered
volume is correct.
If it is a porous package
cover the porous surface.
If it is a burst test reduce
the inflation rate.
For other tests reduce the
test speed.
3203 Flash memory full Not enough space on the
connected USB-stick.
Use an empty USB-stick.
3205 User login not allowed Service access not
enabled.
Enable service access.
(Requires Administrator
access).
3206 Unsupported users
database
You are trying to import
an incompatible user
database
Select a compatible user
database.
3207 Users database corrupted Internal SW error. Try to restart.
3208 Users list import failure Try another USB-stick.
3209 Users list export failure Try another USB-stick.
3215 Cannot edit test definition.
Test definition has been
used
You are trying to edit a test
definition that has been
used.
Either create a new test
definition or change the
“Lock Test setting so that
a used test definition can
be edited.
3216 Test definition value(s)
out of range
Change the value to an
acceptable value.
3236 Test definition exists Save test definition with
another name.
3301 USB stick not inserted No USB-stick connected. Connect USB-stick.
3302 Failed to copy a file to USB Try another USB-stick.
No. Message Problem Solution