34 User Guide EN Dansensor® Lippke® 5000
P/N 380419-C
05/2022
NOTE! Contact your application specialist for correct selection of needles.
A needle is assembled with a hose connector
by pushing the needle into the connector
while turning it approx. 1/3 of a round.
(D) Separate Feed and
Sense
Ideal for very large packages , large packages with multi-
compartment designs and multi-port IV bags.
Recommended for Bubble tests on large packages .
(E) Mono (with Y-piece) Ideal for small packages where there is no room for dual
needles or a bulky test head.
Can be combined with different size needle-heads (incl.
smaller than 4 mm) and hose lengths.
NOTE! Recommended for Bubble tests but not
recommended for fast fill or high accuracy
measurements.