EasyManua.ls Logo

Netafim nmc-pro - EC Sensor Calibration

Netafim nmc-pro
167 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NMC-PRO Installation Manual
148Page
SENSOR INSTALLATION AND DEFINITION
17.1.1 EC Sensor Calibration
1. On EC/pH screen press MENU.
EC Calibration appears.
2. Press Enter. Calib 1.4 appears.
3. Press Enter.
4. Clean and dry an EC sensor. Insert the sensor into 1.4mS calibration buffer; immerse for 10 seconds and
press Enter.
Calibration is in process.
5. When the following screen appears, remove the sensor from the buffer and hold it in the air; press Enter.
6. Calibration is in process; wait until next screen is displayed, which indicates that EC Calibration is complete.
7. Wait until the following screen appears.
EC calibration is complete.
NOTE: If display says "BUFFER FAULT" please refer to Troubleshooting in the EC-pH Manual.
1. On EC/pH screen, press MENU.

Table of Contents

Other manuals for Netafim nmc-pro

Related product manuals