63
Input Configuration Menu
16
ON.L INE
Press ‘RESET/ENTER’ 30) Display flashes "ON.L INE" On line method.
Press ‘䊱/MAX’ 31) Scroll between 2 available selections: "ON.L INE" or "MANUAL" for
the configuration option of your choice. If your choice of selection is
"MANUAL", skip to Step 40).
Press ‘RESET/ENTER’ 32) Display stops flashing and shows "ON.L INE".
Press ‘RESET/ENTER’ 33) Display shows "INPUt 1" Item #1 of coordinate #1.
Pressing ‘RESET/ENTER’ to sequence through
"INPUt 1", Input 1 ’s value
"REAd 1", Read 1 ’s value
"INPUt 2", Input 2 ’s value
"REAd 2", Read 2 ’s value
headings for the desired selection to enter new value.
Press ‘RESET/ENTER’ 34) Display shows "xxxxxx" some 6-digit number, represents the value of
"INPUt 1" as taken from live
input source measurement for desired value.
Press ‘RESET/ENTER’ 35)
Display shows "REAd 1" Item #2 of coordinate #1.
Press ‘RESET/ENTER’ 36) Display shows "xxxxxx" some 6-digit number with first digit flashing,
represents the value of "REAd 1" item #2 of coordinate #1.
Press ‘䊱/MAX’ & ‘
䊳
/MIN’ 37) Use ‘MAX’ and ‘MIN’ buttons to enter new value of "REAd 1".
Press ‘RESET/ENTER’ 38) Display shows "INPUt 2" Item #1 of coordinate #2. Continue to
proceed with the selection and value entry for "INPUt 1" and "REAd 1"
for "INPUt 2" and "REAd 2", (item #1 and #2 of coordinate #2) similar to
steps above.
Press ‘RESET/ENTER’ 39)
Display advances to "MP.SC.OF" Multi-Scale & Offset configuration
menu. If "REAd 2" item #2 of coordinate #2 was prompted or modified, the
controller will also show "StOREd" stored message momentarily.
MANUAL
Press ‘RESET/ENTER’ 40) "MANUAL" heading stops flashing.
Press ‘RESET/ENTER’ 41) Display shows "INPUt 1" Item #1 of coordinate #1.
Pressing ‘RESET/ENTER’ to sequence through
"INPUt 1", Input 1 ’s value
"REAd 1", Read 1 ’s value
"INPUt 2", Input 2 ’s value
"REAd 2", Read 2 ’s value
headings for the desired selection and continue using ‘MAX’ and ‘MIN’
buttons to manually enter or modify the values as described the below
section.
Press ‘RESET/ENTER’ 42) Display shows "xxxxxx" some 6-digit number with first digit flashing,
represents the value of "INPUt 1" item #1 of coordinate #1.
Press ‘䊱/MAX’ & ‘
䊳
/MIN’ 43)
Use ‘MAX’ and ‘MIN’ buttons to enter new value of "INPUt 1" item #1
of coordinate #1.
Press ‘RESET/ENTER’ 44) Display shows "REAd 1". Continue with the sequential items and use
‘MAX’ & ‘MIN’ buttons to select and enter values.
Press ‘RESET/ENTER’ 45) Display advances to "MP.SC.OF" Multi-point Scale & Offset
Configuration Menu. If "REAd 2" item #2 of coordinate #2 was prompted
or modified, the controller will also shows "StOREd" stored message
momentarily.