Multi-Point Scale and Offset Menu Configuration
18
18
ON.L INE IF "ON.L INE" and "NUM.PNt" = 01 are YOUR CHOICE:
(On-Line Calibration with 2 points)
Press ‘RESET/ENTER 6) Display stops flashing "ON.L INE" On-line Calibration or Scaling.
Press ‘RESET/ENTER 7) Display shows "INPUt 0".
Press ‘RESET/ENTER 8) Display shows the actual live-input signal being received and equal to
approximately the minimum input signal or 0% of known load range.
Press ‘RESET/ENTER 9) Display shows "StOREd" momentarily and advances to "REAd 0".
Press ‘RESET/ENTER 10) Display shows last stored value of "REAd 0" with 1st digit flashing.
Press ‘/MAX & ‘
/MIN 11) To enter the value of "REAd 0".
Press ‘RESET/ENTER
This value corresponds to "INPUt 0" in terms of some meaningful
engineering units. To show "INPUt 0" as minimum or 0% as enter
"REAd 0" value = 000000 (Example)
Press ‘RESET/ENTER 12) Display shows "INPUt 1".
Press ‘RESET/ENTER 13) Display shows the 2nd higher live-input signal being connected for
your preference as equal to approximately 100% of the transducer range
or the max. Input signal.
Press ‘RESET/ENTER 14) Display shows "StOREd" momentarily and advances to "REAd 1".
Press ‘RESET/ENTER 15) Display shows last stored value of "REAd 1" with 1st digit flashing.
Press ‘/MAX & ‘
/MIN 16) To enter the value of "REAd 1". This value corresponds to "INPUt 1"
in term of some meaningful engineering units. To show "INPUt 1" as
maximum or 100% as enter "REAd 1" value = 000100 (example).
Press ‘RESET/ENTER 17) Display shows "StOREd" momentarily and advances to "OUt.CNt"
Output configuration menu.
MANUAL
IF "MANUAL" and "NUM.PNt"=01 are YOUR CHOICE
(Scaling without Known Loads or based on specifications)
To scale "MANUAL", calculate inputs based on transducer specifications and manually enter them via
on the front panel push buttons. The following example assumes load cells with this specification:
Maximum Load:....................................................................100.0 lb
Output ................................................................................3.0 mV/V
Sensor Excitation ......................................................................10 V
Maximum Sensor Output = 3.0(mV/V) x 10(V) = 30mV

Table of Contents

Other manuals for Omega DP41-B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omega DP41-B and is the answer not in the manual?

Omega DP41-B Specifications

General IconGeneral
BrandOmega
ModelDP41-B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals