• Off-cart devices connecting to the ultrasound system must comply with the
applicable IEC or national standards, such as IEC 60601-1, IEC 60950, or
the equivalent.
CAUTIONS
• Using accessories, transducers, peripherals, or cables not supplied with the
ultrasound system or recommended by Philips can affect the system in the
form of increased emissions or decreased immunity to external EMI/EMC
occurrences.
• If systems, transducers, and peripherals have been in an environment below
10°C (50°F), allow them to reach room temperature before connecting or
turning them on. Philips recommends allowing 24 hours for complete
normalization. Otherwise, condensation inside the devices could cause
damage. If the device was only briefly exposed to temperatures below 10°C
(50°F), then the time required for the device to return to room temperature
could be significantly less than 24 hours.
NOTE
Any device that is not purchased from Philips or a Philips-authorized agent is
not covered under a Philips service agreement or warranty.
External Printers
You can connect different external printers to your system.
WARNING
Images printed on a report printer are intended only for reference and should
not be used for diagnostic purposes.
NOTES
• Use only the printers listed here with your ultrasound system.
• The system uses the HP Universal Print Driver and the Epson Universal
Print Driver, which support additional printers not listed here. For the
iE33 User Manual
106
4535 616 23901
Preparing the System
4