Immersing Transducers
Immerse this section only, up to 5.1 cm (2 in) from the connector’s strain
relief.
1
This is the maximum allowable immersion depth; you are not required
to immerse to this depth if it is not necessary.
2
Disinfecting TEE Transducers by Immersion
Disinfection by immersion is the accepted method of infection control for
transesophageal transducers. Philips recommends that a protective transducer
cover be used during studies.
311
iE33 User Manual
4535 616 23901
11