CAUTIONS
Before moving the system, ensure that the brakes are completely released
(by fully lifting the brake pedal). Otherwise, the engaged brake pad can
damage the rubber casters on the wheels.
Ensure that the cables for all patient-applied parts are secure before moving
the system. Use the cable management system to ensure that transducer
cables are protected from damage.
Do not roll the system over transducer cables or power cables.
When transporting the system in a vehicle, avoid exposing the monitor to
direct sunlight and do not let the inside temperature of the vehicle exceed
65°C (149°F). Either of these conditions can permanently damage the
monitor.
Preparing and Moving
Follow this procedure to move a system.
1. Press On/Off to turn off the system.
2. When the system shuts down, set Power to off .
3. Disconnect all external cables, including those to power, network, telephone,
and external devices. Secure all cables, transducers, and accessories so that
they do not interfere with the wheels.
4. Retract the keyboard.
5. Squeeze the release lever on the front handle and use the handles to move
the control module to its lowest position.
115
iE33 User Manual
4535 616 23901