TEE Transducer Image Plane Rotation
Calibrating the TEE Transducer
Before connecting the TEE transducer, ensure that the tip and shaft are straight.
When you lock the connector into the transducer receptacle, the TEE transducer
calibrates the array to the 0-degree (horizontal plane) position. As with any array
movement, the motor makes a humming sound while calibrating.
NOTE
Do not turn the TEE control knobs while calibration is in progress.
261
iE33 User Manual
4535 616 23901