If the "Calibration failed" message appears, see "Calibrating the TEE Transducer"
on page 261.
1. Plug the transducer into a transducer receptacle on the system.
2. Turn the latch 90 degrees clockwise to lock the transducer to the system.
Connecting a T6H or T6207 Transducer
You can use a Philips T6H (Omni III TEE) transducer or a Philips T6207 (TEE)
transducer with the iE33 system by using the appropriate adapter:
• For the T6H transducer, use the Omni III adapter
• For the T6207 transducer, use the mini-Multi adapter
These adapters are identical in appearance and operation. Each adapter, however,
supports only the transducer for which it is intended. The Omni III adapter is
shown below. The front of the adapter is the transducer connection. The back
of the adapter is the system connection.
NOTE
Before you connect the T6H transducer, be sure the tip and shaft are straight.
When you connect the transducer, it automatically calibrates itself to the 0-degree
(horizontal plane) position. As with any array movement, the motor makes a
humming sound while calibrating.
1. Plug the Omni III or mini-Multi adapter into the left transducer receptacle on
the iE33 system, with the latch side facing out. (You can use the other
transducer receptacles, but one additional receptacle will be blocked by the
adapter if you do.)
2. Turn the latch on the adapter to lock the adapter to the system.
3. If the "Calibration failed" message appears, see "Calibrating the TEE
Transducer" on page 261.
4. Orient the T6H or T6207 connector so that the cord exits upward, and then
plug it into the adapter.
5. Turn the latch 90 degrees clockwise to lock the transducer into the adapter.
249
iE33 User Manual
4535 616 23901
10
Transesophageal Transducers