Corrective ActionSymptoms
1. Ensure headset is turned on.
2. Ensure headset is paired to system.
3. Refresh the Active Headsets list
and select your headset again.
The system failed to communicate...
error message is displayed during use
of voice control.
Run the system test (see "Testing the
System" on page 343).
An error message is displayed.
1. Click Continue. The system will
power down automatically in
30 minutes.
2. With power off, check for
obstructed air filters (see Cleaning the
System Air Filter).
An error message indicates that the
system is above normal operating
temperature.
Error Messages
The system displays error messages in response to operating or error conditions
detected by the system.
The error messages must be noted and reported to your Philips Ultrasound
customer support representative, who may ask you to run the system test (see
"Testing the System" on page 343).
341
iE33 User Manual
4535 616 23901
12
System Maintenance