Operation of the system without proper and adequate training could lead
to fatal or other serious personal injury.
• Do not operate the system with patients unless you have an adequate
understanding of its capabilities and functions. Using the system without
such understanding may compromise the system's effectiveness and the
safety of the patient, you, and others.
• Never attempt to remove, modify, override, or frustrate any safety device
on the system. Interfering with safety devices could lead to fatal or other
serious personal injury.
• Use the system only for its intended purposes. Do not use the system with
any product that Philips does not recognize as compatible with the system.
Operation of the product for unintended purposes, or with incompatible
products, could lead to fatal or other serious injury.
Electrical Safety
This equipment has been verified by a recognized third-party testing agency as
a Class I device with Type BF and Type CF isolated patient-applied parts, and
Type B non-isolated patient-applied parts. (The safety standards met by this
system are included in the "Specifications" section.) For maximum safety observe
these warnings and cautions:
WARNINGS
• Shock hazards may exist if this system, including all externally mounted
recording and monitoring devices, is not properly grounded. Protection
against electrical shock is provided by grounding the chassis with a
three-wire cable and plug. The system must be plugged into a grounded
outlet. The grounding wire must not be removed or defeated.
• To avoid the risk of electrical shock, never connect the system power cord
to a power strip or an extension cord. When using the power cord, always
connect it directly to a grounded wall outlet.
• Use only Type CF transducers for invasive procedures. Type B and BF
transducers are insufficiently electrically isolated for invasive use.
iE33 User Manual
28
4535 616 23901