S7-3t Lock Operation
Unlocked, freewheeling (white)
Locked (red)
Manipulating the S7-3t TEE Tip
Review the warnings and caution in "Patient Safety During TEE Studies" on page
238 and "TEE Deflection Control Basics" on page 247 before using the transducer
in a study.
1. Put the deflection control knob into freewheeling mode by sliding the
deflection lock so that the indicator is white.
2. Do one of the following:
To deflect the tip toward the posterior, turn the smaller knob clockwise.
To deflect the tip toward the anterior, turn the smaller knob
counterclockwise.
To set the tip in the neutral position (no deflection), turn the smaller knob
so that its white bar is in line with the center of the proximal end of the
transducer handle.
269
iE33 User Manual
4535 616 23901
10
Transesophageal Transducers