To use the temporary ID feature, the Burn Patient Information Into Images
option in the Print/Network setups must be deselected.
When you use the temporary ID workflow, images can be sent to a PACS or
to a DICOM printer before entering actual patient data, if the system is configured
to send or print images as you scan.
NOTES
For an exam started with a temporary ID, edit patient data before ending
the exam. After the exam is ended, you cannot change patient data.
If you have configured a DICOM performed-procedure-step-server, PPS
messages are sent for the temporary ID.
If you have configured a PPS server, PPS discontinue messages are sent for
the temporary ID if you are in send-as-you-scan mode or if you have enabled
send-on-demand.
When the patient data is changed, all images are automatically resent to
any local printers that they were previously sent to.
Starting Emergency Studies
In an emergency, you can use the temporary ID feature to start an exam without
first having to enter patient data.
You can create a temporary ID when starting a protocol, printing, acquiring an
image, or saving a volume.
1. Touch Protocol or press Patient Data, Acquire, or Print, and then click
Temporary ID.
2. When the exam is finished, do either of the following:
To replace the temporary ID with patient data, press Patient Data, click
Edit, and edit the emergency patient data (either manually or by selecting
a Modality Worklist entry from the Worklist tab in the Patient Data
form), and then click End Exam.
To end the exam and use the temporary ID, press End Exam and click
End Exam in the Temporary Patient Study dialog box.
165
iE33 User Manual
4535 616 23901
5
Using the System

Table of Contents

Other manuals for Philips iE33