EasyManua.ls Logo

Philips iE33 - Tip Fold-Over; Recognizing Tip Fold-Over; Correcting Tip Fold-Over

Philips iE33
364 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
transducer can result in lacerations, bleeding, perforation, tearing of
adhesions, and ligament damage. Also be aware that the tip can fold over,
causing similar damage.
Refrain from handling the distal tip whenever possible. If you must handle
the distal tip, grasp it on the sides. Do not touch the top or bottom. Support
the transducer’s proximal head, either by having an assistant hold the
steering mechanism or by clamping the transducer at the steering
mechanism. Ensure that the clamp does not interfere with steering, and do
not clamp any part of the flexible shaft, as this will damage the transducer.
Tip Fold-Over
On rare occasions, the tip of a TEE transducer has folded over during insertion.
The effects can be serious if the situation is handled incorrectly. The esophagus
can be scraped, perforated, or otherwise damaged.
For more information on tip fold-over, see "Tip Fold-Over" in the Help.
Recognizing Tip Fold-Over
The TEE transducer tip might be folded over in the patient if you encounter any
of the following:
Resistance to advancing or removing the transducer
An inability to turn the control knobs
Fixation of the control knobs in the maximum flexion position
Extreme difficulty in obtaining an image
Correcting Tip Fold-Over
If you suspect the transducer tip is folded over, Philips' physician consultants
recommend that you gently try to manipulate the transducer. If the tip is neither
locked nor jammed in a doubled-over position, and you can move it forward,
advance the transducer into the stomach. Then straighten the tip and remove
the transducer.
iE33 User Manual
286
4535 616 23901
Transesophageal Transducers
10

Table of Contents

Other manuals for Philips iE33

Related product manuals