EasyManua.ls Logo

Philips iE33 - Approved Transducers for Electromagnetic Compliance; Approved Accessories for Electromagnetic Compliance

Philips iE33
364 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Approved Transducers for Electromagnetic Compliance
The imaging transducers used with the system may affect its emissions. The
transducers listed in "Clinical Options and Transducers" on page 206, when used
with the system, have been tested to comply with the Group 1, Class B emissions,
as required by International Standard CISPR 11. Use only those transducers.
WARNING
Using cables, transducers, and accessories other than those specified for use
with the system may result in increased emissions from, or decreased immunity
of, the system.
Approved Accessories for Electromagnetic Compliance
Accessories used with the system may affect its emissions. The accessories listed
here, when used with the system, have been tested to comply with the Group 1,
Class B emissions as required by International Standard CISPR 11. Use only the
accessories listed here.
When connecting other accessories to the system, such as a remote video
monitor or computer, it is the user’s responsibility to ensure the electromagnetic
compatibility of the system. Use only CISPR 11 or CISPR 22, Class B-compliant
devices, unless otherwise noted.
WARNING
Using cables, transducers, and accessories other than those specified for use
with the system may result in increased emissions from, or decreased immunity
of, the system.
CAUTION
Use only Class-B USB storage devices with the system. Some plastic-cased
unshielded USB devices may cause RF emissions that exceed Class-B limits. See
the device’s documentation to determine whether it is Class-B compliant.
73
iE33 User Manual
4535 616 23901
2
Safety

Table of Contents

Other manuals for Philips iE33

Related product manuals