EasyManua.ls Logo

Philips iE33 - Control Panel Lateral Translation Force Adjustment

Philips iE33
364 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
e. Switch the lock latch to the right (as shown on the adapter) to lock the
transducer into the adapter.
2. To remove the transducer from the adapter:
a. Switch the lock latch to the left to unlock the transducer from the adapter.
b. Pull the connector toward you.
3. To remove the adapter from the system:
a. Turn the adapter latch to its unlocked position (horizontal).
b. Pull the adapter toward you.
Transducer Maintenance
Transducers require proper care, cleaning, and handling. Reasonable care includes
inspection, cleaning, and disinfection or sterilization, as necessary.
Inspect the transducer cable, case, and lens before each use. Check for cracks
or other damage that jeopardizes the integrity of the transducer. Report any
damage to your Philips Ultrasound customer service representative, and
discontinue use of the transducer.
For all information on transducer cleaning and disinfection and the use of acoustic
coupling gels, see the "Transducer Care" section.
CAUTION
Some ultrasound coupling gels, as well as some solutions for precleaning,
disinfecting, and sterilizing can damage a transducer. Before using a gel or solution
on a transducer, see "Transducer Care". You can also call Philips at 800-722-9377
(North America), or call your local Philips Ultrasound representative (outside
North America).
Information on compatible disinfectants is available on the Philips Ultrasound
Transducer Care website:
www.healthcare.philips.com/us/products/ultrasound/transducers/transducer_care/
If you encounter poor image quality or transducer problems, see
"Troubleshooting" on page 340.
213
iE33 User Manual
4535 616 23901
8
Transducers

Table of Contents

Other manuals for Philips iE33

Related product manuals