Sterilizing Transducers
Sterilization is required if the device is classified as a critical device.
WARNINGS
• Always use protective eyewear and gloves when cleaning, disinfecting, or
sterilizing any equipment.
• In neurosurgical applications, sterilized transducers should be used with a
sterile transducer cover.
• If a premixed solution is used, be sure to observe the solution expiration
date.
CAUTIONS
• Transducers must be cleaned after each use. Cleaning the transducer is an
essential step before effective disinfection or sterilization. Be sure to follow
the manufacturer’s instructions when using disinfectants.
• Use only liquid solutions to sterilize transducers. Using autoclave, gas (EtO),
or other methods not approved by Philips Ultrasound will damage your
transducer and void your warranty.
• Do not allow sharp objects, such as scalpels and cauterizing knives, to touch
transducers or cables.
• When handling a transducer, do not bump the transducer on hard surfaces.
• Ensure the solution strength and duration of contact are appropriate for
sterilization. Be sure to follow the manufacturer’s instructions.
1. Clean the transducer according to "General Cleaning for All Transducers"
on page 302.
2. Mix the sterilization solution compatible with your transducer according to
label instructions for solution strength.
3. Immerse the transducer in the sterilization solution as shown.
4. Follow the instructions on the sterilization label for the duration of transducer
immersion required for sterilization.
iE33 User Manual
318
4535 616 23901
Transducer Care
11