Restricted Use of Isopropyl Alcohol to Clean Transducers
Cable1
Strain relief2
Strain relief/housing joint3
Housing4
2.5 cm (1 in) between housing and strain relief/housing joint5
You can use alcohol in this area of non-TEE transducers6
Do not use alcohol in this area7
1. After cleaning the transducer and cable (see "General Cleaning for All
Transducers" on page 302), wipe or spray the transducer and cable with a
low-level disinfectant. Allow for the manufacturer’s recommended contact
time. For a list of compatible disinfectants, see "Disinfectants Compatibility"
on page 320.
305
iE33 User Manual
4535 616 23901
11
Transducer Care