CAUTIONS
Attempting to disinfect a cable or connector by using a method other than
the one included here can damage the device and voids the warranty.
Orient the parts that must remain dry higher than the wet parts until all
parts are dry.
Use the following procedure to clean the display; the system control panel; all
the external surfaces of the system and the cart; and the ECG trunk cables, leads,
and electrodes.
1. Before cleaning, turn off the system and unplug the power cord from the
power source.
2. Wipe with a soft cloth moistened with soap and water.
CAUTION
Do not spill or spray liquid on the controls, into the system cabinet, or into
the transducer connection receptacle.
3. Remove any solid matter around the keys or the controls with a cotton swab
or toothpick to ensure that solids are not pushed into the cabinet.
4. If blood or other infectious material comes in contact with the system or any
cable other than a transducer cable, wipe with a 70% solution of isopropyl
alcohol.
CAUTION
If blood or other infectious material comes in contact with a transducer or
transducer cable, do not wipe with isopropyl alcohol until you have read the
"Transducer Care" section for specific cleaning guidelines. Isopropyl alcohol
should not be used on some parts of the transducer and should never be
used on any parts of the transducer cable. Additional cleaning agents are also
available for transducers.
5. Remove any residue with a cloth moistened with sterile water.
6. Be sure to dry the equipment to prevent potential corrosion.