EasyManua.ls Logo

Philips iE33 - Page 76

Philips iE33
364 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electromagnetic Immunity: Environment Guidance
Electromagnetic
Environment Guidance
Compliance
Level
IEC 60601
Test Level
Immunity
Test
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Same as
IEC 60601 test
level
± 6 kV contact,
± 8 kV air
Electrostatic
discharge
(ESD),
IEC 61000-4-2
Mains power quality should be
that of a typical commercial or
hospital environment.
Same as
IEC 60601 test
level
± 2 kV for
power supply
lines, ± 8-18 kV
for
input/output
lines
Electrical fast
transient/burst,
IEC 61000-4-4
Mains power quality should be
that of a typical commercial or
hospital environment.
Same as
IEC 60601 test
level
± 1 kV
differential
mode,
±2 kV common
mode
Surge,
IEC 61000-4-5
iE33 User Manual
76
4535 616 23901
Safety
2

Table of Contents

Other manuals for Philips iE33

Related product manuals