3-12 Loading media and calibration
10 Reposition the media by pressing and holding the FORM
FEED/ADVANCE key until the red light no longer shines through the
mark but through the media.
11 Press the SELECT key. The display will show the following prompt:
12 Press and hold the FORM FEED/ADVANCE key until the reflective mark
is positioned behind the Media Sensor Assembly and in close proximity
to the guide arrow indicated on the center plate.
13 Press the SELECT key. If calibration was successful, the following
message will be displayed:
If calibration was unsuccessful, the following message will be displayed:
The user must re-initiate the calibration procedure.
14 After successful calibration, press the PRINT/PAUSE key to return the
printer to a Ready state.
Reflective
Calibration Bad
Reflective
Calibration OK
Place Mark @ Arrow
Then Press Select