27
13.3. Taring
To determine net weight value, load the weighing pan with a packaging, wait
for a stable indication and press
key. Zero indication and the following
pictograms are displayed:
and . The weighing device has been
tared.
Remember not to exceed scale's maximum capacity, i.e. sum of tare weight
value and load weight value must be lower than the maximum capacity value.
Upon unloading of the weighing pan, the sum of tared masses with minus sign
is displayed.
You can assign tare value to a particular product stored in the database.
In such case the assigned tare value is automatically acquired upon selection
of the given product.
It is impossible to tare zero or negative values. When you tare
zero or negative values the scale responds with the following
message:
13.4. Dual Range Devices
N/A in case of single range scales
Switching from weighing with the accuracy of the I weighing range to weighing
with the accuracy of the II weighing range takes place automatically upon
exceeding Max of the I weighing range.
In case of dual range scales:
• upon switching to weighing with the accuracy of the I weighing range,
pictogram/marker is displayed on the left,
• upon switching to weighing with the accuracy of the II weighing range,
pictogram/marker is displayed on the left.
Switching from weighing with the accuracy of the II weighing range
to weighing with the accuracy of the I weighing range takes place
automatically upon unloading of the weighing pan and returning to AUTOZERO
– ( pictogram/marker is displayed).