62
23.1. External Adjustment
External adjustment is carried out using an external mass standard of the right
accuracy and weight value, which value depends on scale type and capacity.
Correction is carried out semi-automatically, successive process stages are
signalled with prompts.
Procedure:
• Enter <Adjustment / External adjustment> submenu. Message
<Remove weight> is displayed.
• Unload the weighing pan and press
key, message <Adjustment;
Please wait...> is displayed.
• Upon completed start mass determination, message <Put weight>
is displayed along with particular value of mass standard assigned to the
scale.
• Load the weighing pan with the required adjustment weight and press
key.
• Upon completed procedure, message <Remove weight> is displayed.
• Take the weight off the weighing pan, the scale displays the
<Adjustment> submenu.
23.2. User Adjustment
User adjustment can be carried out using any adjustment weight, mass
of which ranges between: 0.3 Max - Max. To run user adjustment procedure
enter <Adjustment / User adjustment> submenu.
User adjustment and external adjustment procedures are likewise with one
exception, before user adjustment start, an edit box for entering mass of the
used adjustment weight is opened.
23.3. Start Mass Determination
If the weighing device does not require adjustment or you do not have suitable
amount of adjustment weights, it is possible to determine start mass only.
Procedure:
• Enter <Adjustment / Start mass determination> submenu. Message
<Remove weight> is displayed.
• Unload the weighing pan and press
key, message <Start mass
determination; Please wait...> is displayed.
• Upon completion of start mass determination procedure, the scale
displays the <Adjustment> submenu.