Ethernet
IP Address: IP address of the scale.
Port: number of port for transmission protocol.
Timeout [ms]: time delay - time interval for disconnection, counted from
the moment of the most recent command sent from the peripheral
device to which the weighing device is connected.
*) optional design.
15.2.2. Code Page
In order to provide correct integration of the weighing device with the printer
(correct printout of diacritical signs of a given language) it is necessary to make
sure that the code page of a sent printout is accordant with a code page of
a printer. There are two methods for obtaining code page accordance:
by setting the right code page in the settings of a printer (read the user
manual of the printer) – it must be accordant with the printout code page
of a weighing device:
Code page Language
1250 Polish, Czech, Hungarian.
1252 English, German, French, Spanish, Italian.
1254 Turkish.
by sending the control code from the weighing device, which automatically
sets the right code page of the printer (i.e. code page accordant with the
one of a weighing device).
Default code page value of the printer is 1250 – Central
European code page.
Exemplary weighing device settings for correct printout of Polish signs with use
of EPSON printer connected to RS232 port:
EPSON TM-U220D EPSON TM-T20 EPSON TM-T20
Baud rate 9600 bit/s 38400 bit/s 38400 bit/s
Parity None None None
Code page 852 1250 852
Prefix - 1B742D 1B7412