31.4. Formulation Procedure
• Select the formulation, do it by pressing key to which pictogram
(select formulation) is assigned.
• Save general parameters of 'Formulations' mode to scale memory, the
parameters are to be found upon pressing a key to which
pictogram
is assigned.
• Press key to which
pictogram (process start) is assigned.
When:
1. The declared value of formulation multiplier causes exceeding of the max
capacity value, then the following message is displayed: <Formulation
multiplier value too high>. The process gets cancelled.
2. The declared value of ingredient mass exceeds the max capacity value,
then the following message is displayed: <Ingredient mass value higher
than max capacity value>. The process gets cancelled.
3. Sum of ingredient mass and high deviation values exceeds the max
capacity value, then the following message is displayed: <High deviation
value too high>. The process gets cancelled.
4. The weighing result is unstable then the following message is displayed:
<Unstable weighing result>. The process status changes to PS=Pause
and remains so until stable weighing result is obtained.
• Upon process start, bar graph of weighed ingredient mass is displayed
with the following data:
Top bar pictograms:
process in progress (blinking pictogram).