31.4. Formulation Procedure
Select the formulation, do it by pressing key to which pictogram
(select formulation) is assigned.
Save general parameters of 'Formulations' mode to scale memory, the
parameters are to be found upon pressing a key to which
pictogram
is assigned.
Press key to which
pictogram (process start) is assigned.
When:
1. The declared value of formulation multiplier causes exceeding of the max
capacity value, then the following message is displayed: <Formulation
multiplier value too high>. The process gets cancelled.
2. The declared value of ingredient mass exceeds the max capacity value,
then the following message is displayed: <Ingredient mass value higher
than max capacity value>. The process gets cancelled.
3. Sum of ingredient mass and high deviation values exceeds the max
capacity value, then the following message is displayed: <High deviation
value too high>. The process gets cancelled.
4. The weighing result is unstable then the following message is displayed:
<Unstable weighing result>. The process status changes to PS=Pause
and remains so until stable weighing result is obtained.
Upon process start, bar graph of weighed ingredient mass is displayed
with the following data:
Top bar pictograms:
process in progress (blinking pictogram).