73
Total weight value of all parts loaded onto the weighing pan
cannot be lower than the value determined by „Minimum
reference sample mass” parameter. Unless this condition is
met, the weighing device displays a message:
<Sample mass too low>.
Single part mass must be equal or greater than 0.1 of the
reading unit. Unless this condition is met, the weighing device
displays a message: <Single part mass too low>.
28.5. Setting Reference Sample Mass by Acquiring Mass of a Single Part
from Database
Upon selecting a product from database, single part mass assigned to the
product (<Part mass> entry) is entered automatically.
Procedure:
Enter <Parts counting> mode and press operation panel key to which
pictogram (product database) is assigned, next select given product
from the list.
Weight value of a single part must be declared for the selected
product. To declare weight value of a single part, enter
products database and edit the selected product.
28.6. Entering Reference Sample Mass to Weighing Device Memory
Reference sample mass of single part can be entered to products database.
Procedure:
Press
button, select the given product.
Determine reference sample mass (read sections 28.3 and 28.4 of this
user manual).
Press
button (Assign reference sample), reference sample mass
for the product is recorded in <Part mass> entry.
29. WORKING MODE - PERCENT WEIGHING
Percent weighing is a working mode enabling you to compare the measured
sample with the reference mass. The result is expressed in [%]. Reference
mass can be either determined by weighing or entered to weighing device
memory by an operator.