77
Dosing correction
Enter to correct the dosed mass, this function allows
to optimise the dosing process, change of pressure
of dosed material in the silos is accounted for. Parameters:
Constant correction – specifying global (constant) value
of correction to be applied during each process; Maximum
correction – specifying maximum value of correction, which
can be determined automatically during the process;
Measurements quantity for correction calculation –
specifying how many of the most recent measurements is
to be analysed in order to automatically calculate the
correction during the process.
Readout For detailed description read section 26.1.
Auto threshold For detailed description read section 26.3.
List of manual dosing functions
Min Enter to set the value of low threshold of manual dosing.
Max Enter to set the value of max threshold of manual dosing.
Confirm ingredients manually
Mark to enforce manual confirmation of dosed mass
by means of pressing
key (the function is enabled
for manual dosing mode).
Automatic taring
Enter to enable/disable automatic taring at the moment
of dosing process start.
Readout For detailed description read section 26.1.
Auto threshold For detailed description read section 26.3.
30.3. Automatic Dosing Procedure
• Save general parameters of 'Dosing' mode to scale memory, the
parameters are to be found upon pressing a key to which
pictogram
is assigned.
• In case of selecting a product (that is to be dosed) form the database the
following message is displayed: <Fast dosing threshold, dosing
threshold, dosing outputs, fast dosing outputs will be obtained from
the product data. Continue?> (
- press to select, - press
to cancel).
• Press key to which
pictogram (process start) is assigned, now scale
keypad is locked except for
- stop, - pause, - breakdown keys.