78
When:
1. Dosing outputs are declared incorrectly (e.g. lack of active dosing output)
the following message is displayed: <Dosing outputs determined
incorrectly>, next the process is cancelled.
2. The weighing result is unstable then the following message is displayed:
<Unstable weighing result>. The process status changes to PS=Pause
and remains so until stable weighing result is obtained.
• Upon process start, bar graph of dosed material mass is displayed with
the following data:
Top bar pictograms:
process in progress (blinking pictogram).
Workspace data:
DT1 Fast dosing threshold value.
DT2 Dosing threshold value.
Product Name of a product that is to be dosed.
Tare Tare value (mass of container, silos, etc.).
PS
Process status, the status takes the following values: -3.000kg – amount left to be
dosed; OK – target dosing mass obtained; Pause – process inhibited: a) by
means of
(pause) button, to restart the process press button, b) start signal
is awaited for, c) container empting signal is awaited for; Taring – taring process
in progress; 00:00:05 [s] - automatic countdown of time delay for record of dosed
mass, or automatic countdown of time delay for process end; Chute – pending for
silos empting; Completed – dosing completed; Aborted – dosing aborted.