81
Workspace data:
Min Value of low threshold of manual dosing.
Max Value of max threshold of manual dosing.
PS
Process status, the status takes the following values: -3.000kg – amount left to be
dosed; OK – target dosing mass comprised between MIN and MAX thresholds;
Pause – process inhibited by
(pause) button, to restart the process press
button; Taring – taring process in progress; Completed – dosing completed;
Aborted – dosing aborted.
Product Name of a product that is to be dosed.
Tare Tare value (mass of container, silos, etc.).
Buttons that are assigned to operation panel keys:
Local parameters (button inactive during the ongoing process).
Product selection (button inactive during the ongoing process).
Process start.
Process stop.
Breakdown stop.
• If dosed mass value is between MIN and MAX thresholds then PS=OK
status is displayed.
• In case of active <Confirm ingredients manually> function, the user
must confirm mass value manually by pressing
key.
• Press
key to confirm dosed mass, PS=Completed status
is displayed (process completed). Blinking pictogram
is turned off.
30.5. Dosing Process Report
Dosing report is automatically generated at the end of each dosing process,
next it is sent to port selected for <Peripherals / Printer>. To declare report
content go to <Printouts / Dosing report> submenu. For instruction on how
to declare adjustment report settings read section 'Printouts'.
Report example:
-------------Dosing Report--------------
Start date 2018.07.10 13:21:40
End date 2018.07.10 13:23:28
Operator Nowak Jan
Fast dosing threshold [DT1] 2.800 kg
Dosing threshold [DT2] 3.000 kg
Dosing mass 3.018 kg
Difference 0.018 kg