36
15.2.3. Prefix, Suffix
Prefix and suffix are controlling codes (given in a hexadecimal format) sent
to a printer at print start - <Prefix> parameter, and at print end - <Suffix>
parameter. Sending these codes allows to control globally information and
actions carried out at the beginning and/or at the end of each printout sent from
a weighing device to a printer. Most frequent usage:
• Prefix - sending info on print code page.
• Suffix - sending paper crop command in EPSON printers (in case
of printer equipped with an auto-cutter blade). Paper crop in EPSON
printer – code 1D564108.
Settings of <Prefix> and <Suffix> parameters are valid for all printouts sent via
the weighing device, e.g.: adjustment reports, statistics and header, footer and
GLP printouts.
15.2.4. Record of Measurement Data Onto a USB Flash Drive
Scale program enables record of measurement data onto an external flash
drive.
Procedure:
• Plug a USB flash drive into the USB type A port.
• Enter <Peripherals / Printer / Port> submenu and set <Pendrive> value.
Return to weighing.
• Now the measurement data is saved to a text file on a USB flash drive
whenever you press
key. The file is generated automatically by
a scale software under printout.txt name.
Remove the USB flash drive from the USB port after about
10s counting from the last measurement record, only after
this time the data is saved. Next plug the USB flash drive
to a computer and read the saved text file.
New data is recorded to an existing file therefore it is possible to continue
recording measurement data to the file once created.
USB flash drive must comprise <FAT files system>.