29.4. Reference Sample Mass Entered to the Weighing Device Memory
• Press button (Set part mass), <Set reference sample mass> edit box
featuring an on-screen keyboard is displayed.
• Enter respective value and press
key to confirm changes.
• Now (instead of weight value of the measured load) difference between
values of loaded mass and reference mass is displayed in [%].
29.5. Setting Reference Sample Mass by Acquiring Mass of a Single Part
from Database
Upon selecting a product from database, reference sample mass of single part
assigned to the product (<Reference sample mass> entry) is entered
automatically.
Procedure:
• Enter <Percent weighing> mode and press operation panel key to which
pictogram (products database) is assigned, next select given product
from the list.
• Difference between values of loaded mass and reference mass assigned
to a selected product is displayed in [%].
30. WORKING MODE – DOSING
Dosing is a working mode enabling you to dispense product. Two options are
available, manual and automated dispensing.