Instrument Functions
R&S
®
NRP2
165User Manual 1173.9157.02 ─ 03
Sample Error
Fig. 4-102: Error message / Sample Error
The message indicates that samples were lost while sampling the signal. The measured
value can thus be incorrect.
Trunc. Meas.
Fig. 4-103: Error message / Trunc. Meas.
"Trunc. Meas." is displayed in Burst Av mode, when there is an overflow of the sensor-
internal buffer for samples. Overflow occurs if a burst is too long. The measurement is
aborted.
HW Error
Fig. 4-104: Error message / HW Error
A sensor has detected a hardware error.
Messages and alarms