Remote Control - Commands
R&S
®
NRP2
203User Manual 1173.9157.02 ─ 03
*TRG
TRiGger generates a BUS-trigger event. If the sensor is in the WAIT_FOR_TRG state
and the source for the trigger is set to BUS (TRIG:SOUR BUS), the sensor enters the
MEASURING state. *TRG invalidates all current measuring results. A query of measure-
ment data following *TRG will thus always return the measured value determined in
response to *TRG.
Table 6-3: Error messages:
-211 Trigger ignored: No sensor is in the WAIT_FOR_TRG state.
-211 Trigger ignored: A sensor is in the WAIT_FOR_TRG state but the trigger source is not
BUS.
Usage: Event
*
TST? <error code>
Self TeST query triggers the self test of the instrument and outputs an error code in
decimal form.
Parameters:
<error code> 0 | 1
0
no error
1
an error has occurred
Usage: Query only
*WAI
WAIt to continue prevents the execution of new commands until one of the following
criteria is met:
All previous commands have been executed.
A DCL command is received.
The device is switched off and on again (power-on-reset).
Usage:
Event
6.3 Overview of SCPI Command Systems
The SCPI commands are grouped in command systems according to their function. Only
the most important functions are shown in the figures.
The commands are separated in high-level commands and low-level commands. The
high-level measurement commands combine several low-level setting commands in one
Overview of SCPI Command Systems