Compatibility Information
R&S
®
NRP2
401User Manual 1173.9157.02 ─ 03
D Compatibility Information
This section contains information regarding the compatibility of the R&S NRP2 remote
control commands with those of the "E4418B" and "E4419B" power meters from Agilent.
The R&S NRP2 is largely downward-compatible to the above-mentioned instruments but
offers additional capabilities.
E4418B/E4419B command Compatibility information
CONFigure?
Returns the current configuration. The R&S NRP2 returns the last set-
ting made with CONFigure, as specified by the SCPI standard.
CALCulate:GAIN SENS:CORR:OFFS:STAT
CALCulate:PHOLd
Is implemented in the R&S NRP2 and has the same effect as
CALC:EXTR:RES.
CALibration
:ECONtrol
:RCALibration
:RFACtor
Is not implemented in the R&S NRP2.
DISPlay:ENABle[?]
Is implemented in theR&S NRP2 and has the same effect as
DISPlay:ILLUmination.
DISPlay:SCREen:FORMat
[:NUMeric[1]|2]
Is not implemented in the R&S NRP2.
INITiate
:CONTinuous:SEQUence[1]|2
:IMMediate:SEQUence[1]|2
The :SEQUence node is not implemented in the R&S NRP2.
MEMory:TABLe:GAIN
Agilent specifies the values without a unit; the unit PCT (percent) is
implicitly assumed. The units DB (dB), DPCT (Δ%) and O (unit 1) can
be specified in the R&S NRP2. The default unit is DB.
OUTPut:TRIGger:STATe
Is not implemented in the R&S NRP2.
SENSe
:AVERage:COUNt:AUTO:
SDETect
:DETector:FUNCtion
:MRATe
:SPEed
:V2P
Is not implemented in the R&S NRP2.
SENSe[1|2]:CORRection
:CSET[1]|CSET2[:SELect]
and
SENSe[1|2]:CORRection
:CSET[1]|CSET2:STATe
The R&S NRP2 does not recognize any tables with frequency-depend-
ent calibration values since the sensors have to be calibrated only once
in production. For this reason, there is no compliance with CSET1.
Tables with frequency-dependent offset correction values can be
defined and addressed via :CSET2 (Agilent and R&S NRP2)
or :FDOTable (R&S NRP2 only).
SENSe[1|2]:CORRection:
GAIN[1]?
Has the same effect as SENSe[1..2]:CORRection:CFACtor?