EasyManua.ls Logo

R&S NRP2 - Trigger:all[:Immediate]; Trigger<[1]

R&S NRP2
437 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control - Commands
R&S
®
NRP2
329User Manual 1173.9157.02 ─ 03
A similar response is achieved with the SENSe:BURSt:DTOLerance command in the
Burst mode. However, the Burst mode makes certain that a burst measurement is not
discontinued too early, whereas TRIGger:DTIMe suppresses a new measurement that
is performed too early.
Parameters:
<dtime> float_value
*RST: depending on sensor
TRIGger<[1]..4>:HOLDoff <holdoff>
TRIGger:ALL:HOLDoff <holdoff>
Defines a period after a trigger event within which all further trigger events are ignored.
Parameters:
<holdoff> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit: S
TRIGger<[1]..4>:HYSTeresis <value>
TRIGger:ALL:HYSTeresis <value>
This command is used to specify how far the signal level has to drop below the trigger
level before a new signal edge can be detected as a trigger event. Thus, this command
can be used to eliminate the effects of noise in the signal on the edge detector of the
trigger system.
Unit: DB | PCT
Parameters:
<value> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit: DB
TRIGger<[1]..4>[:IMMediate]
TRIGger:ALL[:IMMediate]
Performs triggering and makes sure that the sensor directly changes from the
WAIT_FOR_TRG state to the MEASURING state irrespective of the selected trigger
source (TRIG:SOUR). A trigger delay set with TRIG:DEL is ignored but not the automatic
delay determined when TRIG:DEL:AUTO:ON is set. When the trigger source is HOLD,
a measurement can only be started with TRIG.
Error messages:
-211 "Trigger ignored": The sensor is not in the WAIT_FOR_TRG state.
Usage: Event
TRIGger

Table of Contents