Remote Control - Commands
R&S
®
NRP2
330User Manual 1173.9157.02 ─ 03
TRIGger<[1]..4>:LEVel <level>
TRIGger:ALL:LEVel <level>
Determines the power a trigger signal must exceed (TRIG:SLOP POS) or fall short of
(TRIG:SLOP NEG) before a trigger event is detected. This setting is only used for the
trigger signal source (TRIG:SOUR INTernal).
Unit: DBM | W
Parameters:
<level> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit: DBM
TRIGger<[1]..4>:
MASTer[:STATe] <state>
Sets the selected sensor in trigger master state. If the trigger master detects an(internal)
trigger event, it generates an external (hardware) trigger signal, which allows the syn-
chronous triggering of other sensors. The trigger master must be not be set to external
triggering, however, all other sensors have too be set to external triggering (TRIG:SOUR
EXT). The trigger signal generated by the trigger master is routed via the connecting cable
to the R&S NRP2 and from there distributed to the other sensors and the trigger output.
Only one sensor can be trigger master at a time.
See also TRIG:SYNC
Parameters:
<state> ON | OFF
*RST: depending on sensor
TRIGger<[1]..4>:SLOPe <state>
TRIGger:ALL:SLOPe <state>
This command determines whether the rising (POSitive) or the falling (NEGative)
edge of the signal is used for triggering.
Parameters:
<state> POSitive | NEGative
*RST: depending on sensor
TRIGger<[1]..4>:SOURce <state>
TRIGger:ALL:SOURce <state>
Sets the trigger signal source for the WAIT_FOR_TRG state.
TRIGger