Remote Control - Commands
R&S
®
NRP2
325User Manual 1173.9157.02 ─ 03
TRIGger:ALL:DELay:AUTO.............................................................................................328
TRIGger<[1]..4>:DTIMe..................................................................................................328
TRIGger:ALL:DTIMe.......................................................................................................328
TRIGger<[1]..4>:HOLDoff...............................................................................................329
TRIGger:ALL:HOLDoff....................................................................................................329
TRIGger<[1]..4>:HYSTeresis...........................................................................................329
TRIGger:ALL:HYSTeresis...............................................................................................329
TRIGger<[1]..4>[:IMMediate]...........................................................................................329
TRIGger:ALL[:IMMediate]...............................................................................................329
TRIGger<[1]..4>:LEVel...................................................................................................330
TRIGger:ALL:LEVel........................................................................................................330
TRIGger<[1]..4>:MASTer[:STATe]...................................................................................330
TRIGger<[1]..4>:SLOPe..................................................................................................330
TRIGger:ALL:SLOPe......................................................................................................330
TRIGger<[1]..4>:SOURce...............................................................................................330
TRIGger:ALL:SOURce...................................................................................................330
TRIGger<[1]..4>:SYNChronize[:STATe]............................................................................331
ABORt<[1]..4>
ABORt:ALL
Immediately sets the respective sensors to the IDLE state. Measurements in progress
are interrupted. If INIT:CONT ON is set, a new measurement is immediately started
since the trigger system is not influenced.
Usage:
Setting only
INITiate<[1]..4>:CONTinuous <state>
INITiate:ALL:CONTinuous <state>
Selects either single-shot or continuous (free-running) measurement cycles. In the
course of a measurement cycle, a sensor passes through the IDLE, INITIATED,
WAIT_FOR_TRG and MEASURING states before it is again set to the IDLE (INIT:CONT
OFF) or the INITIATED (INIT:CONT ON) state. The settings are not changed when
control is switched from manual to remote. INIT:CONT ON is set when remote control
is switched back to manual control.
TRIGger