Remote Control Fundamentals
R&S
®
NRP2
183User Manual 1173.9157.02 ─ 03
SCPI command Value Remark
TRIGger[1..4]:SLOPe
- Unchanged, is irrelevant.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual mode.
TRIGger[1..4]:DELay
- Unchanged, is irrelevant.
TRIGger[1..4]:HOLDoff
Depends on sensor User-selectable.
TRIGger[1..4]:LEVel
Depends on sensor User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:
STARt
Depends on sensor User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:STOP
Depends on sensor User-selectable.
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual mode.
Mode Timegate Av
TRIGger[1..4]:SOURce EXTernal |
INTernal
User-selectable.
TRIGger[1..4]:SLOPe POSitive |
NEGative
User-selectable.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual mode.
TRIGger[1..4]:DELay
Depends on sensor User-selectable.
TRIGger[1..4]:HOLDoff
Depends on sensor User-selectable.
TRIGger[1..4]:LEVel
Depends on sensor User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:
STARt
- Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:STOP
- Unchanged, is irrelevant.
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual mode.
Mode Timeslot Av
TRIGger[1..4]:SOURce EXTernal |
INTernal
User-selectable.
TRIGger[1..4]:SLOPe POSitive |
NEGative
User-selectable.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual mode.
TRIGger[1..4]:DELay
Depends on sensor User-selectable.
TRIGger[1..4]:HOLDoff
Depends on sensor User-selectable.
TRIGger[1..4]:LEVel
Depends on sensor User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:
STARt
- Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:STOP
- Unchanged, is irrelevant.
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual mode.
Mode Trace & Statistics
Automatic Settings for Manual Selection of a Measurement Mode