Remote Control - Commands
R&S
®
NRP2
233User Manual 1173.9157.02 ─ 03
CALCulate<[1]..4>:STATistics:MARKer: HORIzontal:POSition[:X][:POWer]
<power>
Is available with statistics measurements (CCDF and PDF). It sets the horizontally shift-
able marker on the power axis. The absolute position is specified. The statistics measured
value can then be read out with CALC:STAT:MARK:HOR:DATA? if the measurement at
this position has a valid value.
Parameters:
<power> float_value, units: DBM | DBUV | W
Default unit: DBM
CALCulate<[1]..4>:STATistics:MARKer:HORizontal:DATA?
Is available with statistics measurements (CCDF and PDF). It returns the value of the
statistics function at the position of the marker if the value is valid.
Unit: O
Usage:
Query only
CALCulate<[1]..4>:STATistics:MARKer:VERTical:POSition[:Y] <position>
Is available with statistics measurements (CCDF and PDF). It sets the vertically shiftable
marker on the statistics axis. The associated value on the power axis can then be read
out with CALC:STAT:MARK:VERT:DATA? if such a value is available.
Unit: O
Parameters:
<position> float_value
CALCulate<[1]..4>:STATistics:MARKer:VERTical:DATA[:POWer]?
Is available with statistics measurements (CCDF and PDF). It returns the absolute posi-
tion of the marker on the power axis.
Unit: DBM
CALCulate (Configuration of Calculate Blocks)