Remote Control - Commands
R&S
®
NRP2
262User Manual 1173.9157.02 ─ 03
Parameters:
<mode> NONE
No additional values are measured.
MINMax
By averaging the measured values in the sensors, extreme values
are lost. Using [SENSe<[1]..4>]:AUXiliary MINMax , a
sensor can be assigned to transmit to the R&S NRP2 the minima
and maxima that occurred in the Trace mode together with the
measured value. If CALCulate:FEED "POWer:PEAK ON
SWEep<[1]..4>" is set, peak (instead of average) values are
used for calculating.
RNDMax
In contrast to MINMax, instead of the Min value the value of a ran-
domly selected sample is returned. All evaluations occur using
these values instead of the average values.
*RST:
depending on sensor
[
SENSe<[1]..4>]:AVERage:COUNt <filter_length>
Sets the filter length i.e. the number of readings to be averaged for one measured value..
The higher the count the lower the noise and the longer it takes to obtain a measured
value.
Parameters:
<filter_length> int_value
Range: depending on sensor
*RST: depending on sensor
Default unit: -
[SENSe<[1]..4>]:AVERage:COUNt:AUTO <state>
Can be used to automatically determine a value for the filter legth
[SENSe<[1]..4>]:AVERage:COUNt. If the automatic switchover is activated with the
ON parameter, the sensor always defines a suitable filter length which can also be quer-
ied via SENS<[1]..4>:AVER:COUN?.