Remote Control - Commands
R&S
®
NRP2
273User Manual 1173.9157.02 ─ 03
[SENSe<[1]..4>][:POWer]:TGATe[:EXCLude]:MID[:STATe] <state>
Activates/deactivates the blanking out of time intervals in the timegates, affects all gates.
Parameters:
<state> ON | OFF
*RST: OFF
[SENSe<[1]..4>][:POWer]:TGATe<[1]..4>[:EXCLude]:MID:TIME <time_interval>
Together with SENS:TGAT:MID:OFFS, allows a time interval in a timegate to be blanked
out. SENS:TGAT:MID:TIME determines the distance from the start of the measurement
to the start of the interval to be blanked out (see figure below).
Trigger level
Power
Time
Trigger event
Gate 4
SENS:TGAT2:OFFS:TIME
SENS:TGAT2:TIME
SENS:TGAT2:MID:TIME
SENS:TGAT2:MID:OFFS:TIME
Fig. 6-11: MID parameter in the Timegate mode
Parameters:
<time interval> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit: S
[SENSe<[1]..4>][:POWer]:TGATe<[1]..4>:OFFSet:TIME <time>
Is used for configuring the Timegate mode. It specifies the start of one of the four gates
(sweeps) relative to the delayed trigger point. If this value is negative, the gate starts
before the trigger point. This is possible because the sensor stores samples of the mea-
sured powers and can thus transfer the start of a measurement to the past. In manual
operation, the start and duration of the gates can also be defined graphically if the sensor
is operated in the Trace mode.
SENSe (Sensor Commands)