Remote Control - Commands
R&S
®
NRP2
291User Manual 1173.9157.02 ─ 03
SERVice<[1]..4>:RESult.................................................................................................292
SERVice<[1]..4>:SERRors..............................................................................................292
SERVice<[1]..4>:UNLock................................................................................................292
SERVice<[1]..4>:CALibration:TEST? <NR1>
Tests the consistency of the calibration data set according to three criteria:
If the following bit is set in
<nr1>,
the following action is executed.
0 The serial number of the calibration data set is tested.
1 The checksum of the calibration data set is tested.
2 The header of the calibration data set is tested.
If an inconsistency of the data set is detected, the query returns a 1, otherwise a 0.
Parameters:
<NR1> int_value
Range: 0 to 255
Usage: Query only
SERVice<[1]..4>:CALibration:TEMP <state>
Initiates a temperature measurement of the sensor in question. The temperature can be
queried with SERV:CAL:TEMP:DATA?.
Parameters:
<state> ONCE
Usage: Event
SERVice<[1]..4>:CALibration:TEMP:DATA?
Returns the temperature of the sensor in Kelvin if a temperature measurement was pre-
viously initiated by means of SERV:CAL:TEMP ONCE.
Usage:
Query only
SERVice:DISPlay:OVERload <state>
Switches the sensor overload warning on/off.
Attention:
If the sensor is overloaded, it could be damaged! The warning should not be suppressed
if you are not certain that the sensor is out of danger.
Parameters:
<state> ON | OFF
*RST: ON
SERVice