Remote Control - Commands
R&S
®
NRP2
311User Manual 1173.9157.02 ─ 03
Bit No. Meaning
3 Window 3 Power
The measured values returned by Calculate block 3 are corrupt.
4 Window 4 Power
The measured values returned by Calculate block 4 are corrupt.
5 Window 5 Power
The measured values returned by Calculate block 5 are corrupt.
6 Window 6 Power
The measured values returned by Calculate block 6 are corrupt.
7 Window 7 Power
The measured values returned by Calculate block 7 are corrupt.
8 Window 8 Power
The measured values returned by Calculate block 8 are corrupt.
9 to 14 Not used
15 Bit 15 will never be used.
A bit is set if the error -231, "Data questionable", occurs in the respective calculate block.
6.13.19 Questionable Calibration Status Register
The EVENt register and the CONDition register contain information as to whether the
zero offset of a sensor is still valid.
The Questionable Calibration Status Register can be read by the following commands:
STATus:QUEStionable:CALibration[:SUMMary]:CONDition?
STATus:QUEStionable:CALibration[:SUMMary][:EVENt]?
Table 6-31: Meaning of bits used in the Questionable Calibration Status Register
Bit No. Meaning
0 Not used
1 Sensor A Calibration
Zeroing of sensor A was not successful.
2 Sensor B Calibration
Zeroing of sensor B was not successful.
3 Sensor C Calibration
Zeroing of sensor C was not successful.
4 Sensor D Calibration
Zeroing of sensor D was not successful.