Programming Measurements
R&S
®
NRP2
399User Manual 1173.9157.02 ─ 03
SENS:FUNC "POW:AVG" 'ContAv mode
SENS:BUFF:STAT OFF 'no buffered measurements
SENS:APER MIN 'integration time as short as possible
SENS:RANG:AUTO OFF 'automatic range selection OFF
SENS:CORR:OFFS:STAT OFF 'no global offset correction
SENS:CORR:DCYC:STAT OFF 'no duty cycle correction
SENS:CORR:FDOT:STAT OFF 'no frequency-dependent offset correction
SENS:SGAM:CORR:STAT OFF 'no measured-value correction for
'reflection coefficient of source
TRIG:DEL 0 'no delay on triggering
TRIG:DEL:AUTO OFF 'no delay on triggering
CALC1:MATH "(SENS1)" 'only uses the first CALCulate block
CALC2:MATH "(SENS2)"
CALC3:MATH "(SENS2)"
CALC4:MATH "(SENS2)"
CALC5:MATH "(SENS2)"
CALC6:MATH "(SENS2)"
CALC7:MATH "(SENS2)"
CALC8:MATH "(SENS2)"
Optimization of measurement speed

Table of Contents

Related product manuals