Cubis MSA User Manual 119
Checkweighing
Execute Task: Checkweighing
D t If you haven’t already done so, go to Application Management.
y The task selection is displayed.
t Touch the desired task.
or
t If the desired task is already selected, touch 5VCTV.
Define check values by entry:
t Touch the input field Setp and enter the target.
t Touch the input field Min or Max and enter the minimum or maximum check
values.
or
Define check values with current load:
t Place the weight on the weighing pan.
t Touch /QTG.
t Touch the field of the check value whose current weight value you want to use.
t If required, repeat the process with additional weights for the other check values.
t To exit the input mode without saving the value, touch $CEM.
t To begin checkweighing, touch 5VCTV.
y The entered check values are displayed graphically.
t Place the object to be checked on the weighing pan.