62 Cubis MSA User Manual
Task Management
Standard printout parameters when the balance is first delivered
When the balance is first delivered, two interfaces are configured: The peripheral
port (Com A) is configured for one printout, the US interface (Com B) is configured
for data with a PC per SBI protocol.
Interfaces: Com A (peripherals port) and Com B (US/PC)
Protocol for Com A: Print (pre-set by the interface configuration)
Standard for output to Com A: Standard 1 (measured value only)
GLP-compliant printout to Com A: Off
Print event for output to Com A: Press the PRINT Pkey
Element to be output via Com A: Measured value
Protocol to Com B: SBI (pre-set by the interface configuration)
SBI output to Com B: Single printout to the PC
Criterion for output to Com B: After stability of balance
Format for output to Com B: 22 characters
Tare after printing: Off
t You can select the variable parameters individually by clicking on the respective
input field.
or
t To be guided step-by-step through the settings, use the wizard.
Example: Manual printout to peripherals port, SBI operating mode
y After manual operation, you want an individual printout to be printed to the
ported peripheral device.
Interface: Com A (peripherals port)
Protocol: SBI (toggling to the interface configuration possible)
SBI Printout: Individual printout
Criterion: Manual without stability
Format: 22 characters (for applications)
Tare after printing: Turned on or off
Example: Automatic printout to peripherals port, SBI operating mode
y At a pre-set interval, automatic printouts should be sent to the ported
peripheral device.
Interface: Com A (peripherals port)
Protocol: SBI (toggling to the interface configuration possible)
SBI Printout: Automatic printout
Criterion: Without stability, at stability or after load change
Format: 22 characters (for applications)
Stop automatic printing: Not possible or by using the Print Pkey.
Interval for automatic printing: Number of cycles (1, 2, 5, 10 to100) and 0.1, 0.2,
0.5 to 10 seconds.