Cubis MSA User Manual 63
Task Management
Example: Automatic printout to peripherals port; remote display operating mode
y At a pre-set interval, automatic printouts should be sent to the ported
peripheral device as remote display.
Interface: Com A (peripherals port)
Protocol: Remote display (toggling to the interface configuration possible)
SBI Printout: Automatic printout
Criterion: Without stability, at stability or after load change
Format: 22 characters (for applications)
Interval for automatic printing: Display updates (1, 2, 5, 10 to100) or every 0.1, 0.2,
0.5 to 10 seconds.
Example: Printout to peripherals port only by using the PRINT P key; printer operating
mode
y After pressing the Print key, the measured value should be printed to the ported
peripheral device.
Interface: Com A (peripherals port)
Protocol: Print (toggling to the interface configuration possible)
Standard data output: Standard 1 (measured value only), Standard 2 (measured
value and date/time), Standard 3 (measured value with date/time and weight block)
GLP-compliant printout Turned on or off
Print event: Press the PRINT key
Elements for the Print event PRINT: Measured value PRINT.MEASURE
Tare after printing: Turned on or off
Setting Options of Printout Parameters
Standard Data Output
You can select which data should be printed out as default.
5VCPFCTF (measured value only)
5VCPFCTF (measured value with date/time)
5VCPFCTF (measured value with date/time and weight block)
7UGTKPRWV is only selectable if the standard settings have been modified by
the user.
GLP-compliant Printout
You can select if and when the printout should be GLP-compliant
1HHPQ).2RTKPVQWV
(QTECNKDTCVKQPCFLWUVOGPVQPN[
#EVKXCVGFKPVJGCRRNKECVKQP
#NYC[UQP
7UGTKPRWV
is only selectable as long as none of the three options apply:
i.e. if only a GLP header but not a GLP footer has been configured.